- 5 -
electrical calibration of the unit.
The LCD displays "CAL" for about 1
second. As needed, use the calibration
tool to adjust the Calibration (CAL) VR
until "94.0dB" is displayed.
This ends the calibration. Make sure
that the Measurement Slide Switch is
returned to the "MEAS" (up) position
and the Battery Cover is reattached.
2. Acoustic calibration using a standard
sound calibrator such as Sper Scien-
tific's 840031.
With the Measurement Slide Switch in
the "MEAS" (up) position, place the mi-
crophone in the calibrator.
As needed, use the calibration tool to
adjust the Calibration (CAL) VR until
"94.0dB" is displayed.
Measurement
1. Press the Power / Max Hold button to
turn the meter on.
2. Point the Microphone at the noise
source.
3. The LCD displays the measured sound
level. Readings are updated twice
every second.
"OVER" is displayed when the sound
Przeglądanie stron 4